Decreasing Load Test

The Decreasing Load Test checks the scale accuracy when incrementally removing weight from the Top Plate. Use weights that correspond to the NCR 7895 programmed unit of weight.

Note

This test must immediately follow the Over-Capacity Test. Do NOT remove any of the weights from the previous test prior to running this test.

For Metric Scales, kg
Step Applied LoadScale Display ResultTolerance
115.000 kg15.000± 0.0075 kg
210.000 kg10.000± 0.005 kg
35.000 kg5.000± 0.005 kg
42.500 kg2.500± 0.0025 kg
50.100 kg0.100± 0.0025 kg
For US Scales, lb
StepApplied LoadScale Display ResultTolerance
130.00 lb 30.00± 0.015 lb
220.00 lb 20.00 ± 0.01 lb
310.00 lb 10.00± 0.01 lb
45.00 lb5.00± 0.005 lb
50.10 lb 0.10± 0.005 lb